Huge MTF error for sample data

2 replies [Last post]
nihy
Offline
Joined: 1 Dec 2018

Hi, I am a new user of FindDQE. When I was running sample data, I found a huge difference of MTF compared to provided result. When I was testing DE12, it even gave a DQE near 3.0, so I would like to ask what can be the source of this error? The parameter I input was the same to the .com file provided.

Thanks in advance.

Haoyang

niko
Offline
Joined: 5 May 2009
The exposure you entered

The exposure you entered could be incorrect. How did you calibrate the exposure?

nihy
Offline
Joined: 1 Dec 2018
As in your script, I chose

As in your script, I chose total dose(2) and input 52.8 for total electron dose/pixel.

The following is a copy of my input in bash. What puzzles me is in the reference paper, it gives that DE12 data collection with 13 eps and 25 frames per second, so I wonders how 52.8 total dose/pixel comes from.

You can input either the counts/electron (gain conversion factor) or the total electron dose/pixel.
Enter 1 to input counts/electron, 2 for total dose/pixel
2
2

Enter total electron dose/pixel
52.8
52.7999992

Enter counts multiplier (usually 1).
1
1.00000000

Enter number of images
(usually 1; 2 if also giving flat field image)
1
1

Input image with pointer silhouette
DE12-200kv-50e-13eps.mrc
DE12-200kv-50e-13eps.mrc